(b)(4).Date sent: 5/12/2026 d4: batch # 354d35.Investigation summary the product was returned to ethicon endo surgery for evaluation.Visual inspection and functional testing were conducted on the returned device.Visual analysis of the returned sample determined that the glc80 device was received with no apparent damage and with a glcr80g reload loaded in the device.The reload had the proximal 34 drivers up without staples, and the remaining drivers down with staples present.The returned reload had the swing tab in the locked position.Further evaluation shows nicks in the knife.One possible cause for this type of damage to the knife is when the device is fired over an already existing staple line, hard object or thicker tissue than indicated.The reload was reset and the device was tested for functionality with the returned reload and it fired, cut, and formed all the remaining staples as intended.The staple line and cut line were complete and the staples meet the staple form release criteria.The event reported was confirmed and it is related to improper use of the device.It should be noted that the cartridge reload is designed to lockout, as a safety feature, if any staples have been fired from the cartridge reload.In addition, failure to complete the stroke may result in incomplete staple line.Please reference the instruction for use for more information.As part of ethicon endo surgery¿s quality process all devices are manufactured, inspected, and released to approved specifications.Device history review a manufacturing record evaluation was performed for the finished device batch number 354d35, and no non-conformances were identified.This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by ethicon, or its employees that the report constitutes an admission that the product, ethicon, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
|
|
It was reported that during colorectal procedure the surgeon attempted to staple the incision; however, the stapler did not cut completely through the tissue, leaving approximately 2 cm uncut at the end of the staple line.There were no patient consequences or adverse outcomes.A new stapler was obtained, and the surgeon successfully completed the case.
|