| Model Number |
NT821707 |
| Medical Device Problem Code |
Material Split, Cut or Torn (4008)
|
| Health Effect - Clinical Codes |
Foreign Body In Patient (2687); No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
04/16/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
Part nt821707 - the tip of a lumber catheter sheared while placing in a patient.
|
| |
|
Additional Manufacturer Narrative
|
|
Initial report ref to natus complaint# (b)(4).Per (b)(4) rev 06 natus external drainage lumbar catheters - risk analysis spreadsheet hazard 1.1 - catheter tubing is cut by sharp inner edge of needle while accessing into lumbar spine and a catheter piece is left in the patient body.Effect (harm): adverse tissue reaction residual risk: moderate the hazard identified have been reduced as far as possible, and the residual risk for these hazards is mainly associated with patient having a adverese tissue reaction.The device is to be used in sterile condition and the ifu contains information about proper usage.In addition, techniques limiting cross contamination are already in place in the environment for use.The customer was provided with product return instructions.Further investigation to be carried out.
|
| |
|