• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: NATUS MEDICAL INCORPORATED NT821707, DRAINAGE KIT, LUMBAR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

NATUS MEDICAL INCORPORATED NT821707, DRAINAGE KIT, LUMBAR Back to Search Results
Model Number NT821707
Medical Device Problem Code Material Split, Cut or Torn (4008)
Health Effect - Clinical Codes Foreign Body In Patient (2687); No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 04/16/2026
Type of Reportable Event Malfunction
Event or Problem Description
Part nt821707 - the tip of a lumber catheter sheared while placing in a patient.
 
Additional Manufacturer Narrative
Initial report ref to natus complaint# (b)(4).Per (b)(4) rev 06 natus external drainage lumbar catheters - risk analysis spreadsheet hazard 1.1 - catheter tubing is cut by sharp inner edge of needle while accessing into lumbar spine and a catheter piece is left in the patient body.Effect (harm): adverse tissue reaction residual risk: moderate the hazard identified have been reduced as far as possible, and the residual risk for these hazards is mainly associated with patient having a adverese tissue reaction.The device is to be used in sterile condition and the ifu contains information about proper usage.In addition, techniques limiting cross contamination are already in place in the environment for use.The customer was provided with product return instructions.Further investigation to be carried out.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
NT821707, DRAINAGE KIT, LUMBAR
Common Device Name
NT821707, DRAINAGE KIT, LUMBAR
Manufacturer (Section D)
NATUS MEDICAL INCORPORATED
5955 pacific center boulevard
san diego CA 92121
Manufacturer Contact
5955 pacific center boulevard
san diego, CA 92121
MDR Report Key25157950
Report Number2023988-2026-00059
Device Sequence Number9581987
Product Code JXG
UDI-Device Identifier00382830050531
UDI-Public00382830050531
Combination Product (Y/N)N
Initial Reporter CountryUS
PMA/510(K) Number
K964923
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Other
Type of Report Initial,Followup,Followup
Report Date (Section B) 06/14/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Model NumberNT821707
Device Catalogue NumberNT821707
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? No
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 04/16/2026
Supplement Date Received by ManufacturerNot provided
Not provided
Initial Report FDA Received Date05/13/2026
Supplement Report FDA Received Date06/10/2026
06/14/2026
Was Device Evaluated by Manufacturer? (Y/N) Yes
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient SexUnknown
-
-