| Model Number |
RE-1021 |
| Medical Device Problem Code |
Fluid/Blood Leak (1250)
|
| Health Effect - Clinical Codes |
Insufficient Information (4580); No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
03/26/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
The surgeon reported the lcs leak during the implantation.No consequence of the patient has been reported, the device has been replaced during the surgery.
|
| |
|
Event or Problem Description
|
|
The surgeon reported the lcs leak during the implantation.No consequence has been reported, the device has been replaced
during the surgery.
|
| |
|
Additional Manufacturer Narrative
|
|
Investigation of the returned device revealed a deep cut on the reservoir guard, resulting in material tearing and leakage.The observed damage is consistent with improper handling/manipulation of the device.No manufacturing deficiency was identified.
|
| |
|