|
This supplemental report is being submitted to provide the results of the final investigation.Updated fields: d8, d9, g3, g6, h2, h3, h4, h6 and h11.The device was returned to olympus for inspection, and the reported failure was confirmed.In addition, the following reportable malfunctions were identified during the device evaluation: the cord was found to be completely broken at the plug (male connector) for the electrosurgical generator.Charred stains were observed around the damaged area, consistent with thermal damage.Based on the results of the investigation a root cause could not be identified.A review of the device history record found no deviations that could have caused or contributed to the reported issue.Should additional relevant information become available, a supplemental report will be submitted.Olympus will continue to monitor field performance for this device.
|