| Model Number |
694765 |
| Medical Device Problem Codes |
Signal Artifact/Noise (1036); Over-Sensing (1438)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
04/20/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that the right ventricular (rv) lead triggered a lead noise warning and a lead integrity alert (lia) due to high rate non-sustained episodes and the sensing integrity counter (sic) being incremented.The rv lead also exhibited oversensing before and after detection, resulting in earlier classification of episode type and oversensing before detection, resulting in misclassification of episode type.The lead remains in use.No patient complications have been reported as a result of this event.
|
| |
|
Additional Manufacturer Narrative
|
|
Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
| |
|