Device evaluation details: the device was returned to johnson & johnson medtech (j&j medtech) for evaluation.Following j&j medtech procedures, a visual inspection and magnetic test of the returned device were performed.Visual inspection reddish material and a hole in the surface of the pebax.The device was connected to the carto 3 system and it was recognized and visualized correctly.No magnetic issues or errors were observed.A manufacturing record evaluation was performed for the finished device batch number, and no internal actions were identified.The reddish material inside the pebax could be related to the magnetic sensor error reported by the customer; therefore, the customer complaint was confirmed.The potential cause of the pebax damage could be related to the manipulation of the device during the procedure; however, this cannot be conclusively determined.The instructions for use (ifu) contain the following recommendations: do not scrub or twist the tip electrode because damage to the tip electrode bond may occur and loosen the tip electrode, or damage the contact force sensor and affect measurement accuracy.In addition, in order to prevent damage to the catheter tip, verify compatibility between the sheath and catheter, advance the catheter through the sheath prior to insertion.Any sheath < 8.5 f is contraindicated.Regarding the magnetic sensor error reported by the customer, the carto® 3 system instructions for use (ifu) contain the following information: the magnetic sensor of the catheter is disconnected.If the problem persists, replace the catheter cable or the catheter.As part of johnson & johnson medtech¿s quality process, all devices are manufactured, inspected, and released to approved specifications.This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by biosense webster inc., or its employees that the report constitutes an admission that the product, biosense webster inc., or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.Manufacturer's ref.# (b)(4).
|
|
It was reported that a patient underwent a paroxysmal atrial fibrillation (afib) ablation procedure with a thermocool smarttouch sf catheter for which biosense webster¿s product analysis lab (pal) identified a hole on the pebax.It was initially reported by the customer that catheter icon was moving erratically during rf delivery only.There was also an unknown, intermittent magnetic distortion error present on the carto 3 system.To troubleshoot they exchanged the indifferent electrodes, exchanged the cable to the catheter, without resolution.The catheter was exchanged to resolve the error and continue the case successfully.There was no patient consequence.The customer's reported sensor and visualization issues is not considered to be mdr reportable since the potential risk that it could cause or contribute to a serious injury or death to the operator or patient is remote.On 17-apr-2026, the bwi pal revealed that a visual inspection of the returned device found a hole in the pebax and a reddish material inside the pebax.These findings were reviewed and assessed the issue of a ¿hole¿ in the pebax as an mdr reportable malfunction since the integrity of the device has been compromised.
|