BECTON DICKINSON & CO (FRANKLIN LAKES) BD VACUTAINER® SODIUM FLUORIDE/POTASSIUM OXALATE 10MG/8MG; TUBES, VIALS, SYSTEMS, SERUM SEPARATORS, BLOOD COLLECTION
|
Back to Search Results |
|
| Catalog Number |
367922 |
| Medical Device Problem Code |
Short Fill (1575)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
03/19/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
E.1.Initial reporter facility name: (b)(6) hospital.Investigation summary: bd did not receive any samples or photos for investigation.A total of 100 retained samples were visually inspected, and the stoppers were correctly placed on the tubes.Functional tests were performed on 10 retained samples.All tubes were within specification limits, and no issues were identified.Based on a review of the device history record for the incident lot, all product specifications and requirements for lot release were met.This complaint has not been confirmed for the indicated failure mode: underfill.Complaints received for this device and reported condition will continue to be tracked and trended.Information will be captured on trend reports and monitored.Our business team regularly reviews the collected data for the identification of emerging trends.
|
| |
|
Event or Problem Description
|
|
It was reported while using bd vacutainer® sodium fluoride/potassium oxalate 10mg/8mg, an unspecified number of tubes underfilled.No adverse impact was reported regarding the patient or user.
|
| |
|
Search Alerts/Recalls
|
|
|