Model Number 0174
Medical Device Problem Code High impedance (1291)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 04/22/2026
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that this cardiac resynchronization therapy defibrillator (crt-d) system exhibited an abrupt high out of range shock impedance measurement.The patient experienced a fall, which technical services (ts) suspected caused damage to the system.An evaluation was performed with vector breakdowns and isometrics, and all measurements were within range.Ts recommended programming options.No adverse patient effects were reported.This right ventricular (rv) lead remains in service.
 
Additional Manufacturer Narrative
This lead was not returned for analysis, as it remains implanted and in service; therefore, a technical analysis could not be performed.Based on a thorough review of the reported complaint, boston scientific has assigned an investigation conclusion code of known inherent risk of device.If pertinent information is provided in the future, a supplemental report will be submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ENDOTAK RELIANCE G
Common Device Name
IMPLANTABLE LEAD
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
4100 hamline avenue north
saint paul MN 55112
Manufacturer (Section G)
CPI - DEL CARIBE
guidant puerto rico b. v.
#12, rd. #698
dorado PR 00646
Manufacturer Contact
timothy degroot
4100 hamline avenue north
saint paul, MN 55112
6515826168
MDR Report Key25164107
Report Number2124215-2026-25913
Device Sequence Number10737360
Product Code LWS
UDI-Device Identifier00802526373640
UDI-Public(01)00802526373640(17)070423
Combination Product (Y/N)Y
Initial Reporter StateMA
Initial Reporter CountryUS
PMA/510(K) Number
P910073
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2005
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Company Representative
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date (Section B) 05/13/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Expiration Date04/23/2007
Device Model Number0174
Device Catalogue Number0174
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 04/22/2026
Initial Report FDA Received Date05/13/2026
Date Device Manufactured04/23/2005
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient Age78 YR
Patient SexMale
-
-