| Catalog Number |
00514010400 |
| Medical Device Problem Codes |
Device Emits Odor (1425); Sparking (2595)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
10/23/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that during cleaning the device showed sparks and had a burning smell.There was no patient harm/injury or surgical delay as there was no patient involvement.Due diligence is complete and there is no additional information available.There was no adverse event associated with this malfunction.
|
| |
|
Additional Manufacturer Narrative
|
|
This complaint is recorded by zimmer biomet under (b)(4).Review of the most recent repair record determined the burning smell issue was confirmed as cart¿s motherboard display terminal had burn damage.Replaced the cart motherboard and resolved the reported issue.Review of the device history record identified no deviations or anomalies during manufacturing.A definitive root cause cannot be determined.The reported event was confirmed.If any further information is found which would change or alter any conclusions or information, a supplemental report will be filed accordingly.Zimmer biomet will continue to monitor for trends.
|
| |
|