• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ETHICON INC. STRATAFIX; SUTURE, SURGICAL, ABSORBABLE, POLYDIOXANONE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ETHICON INC. STRATAFIX; SUTURE, SURGICAL, ABSORBABLE, POLYDIOXANONE Back to Search Results
Catalog Number SXPP2B405
Medical Device Problem Code Material Separation (1562)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 01/01/2026
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that a patient underwent an unknown procedure in 2026 and barbed suture was used.During the procedure, it was reported by the sales rep that as the orthopedic surgeon was closing in his total joint cases, "the needles has been popping off the suture all day." the materials coordinator shared, "this has resulted in the circulator needing to open free needles for him to proceed closing." there was no patient consequence reported.The device was returning.No adverse patient consequences were reported.Additional information was requested.
 
Additional Manufacturer Narrative
Product complaint # (b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803, part 4 subpart b.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by ethicon, or its employees that the report constitutes an admission that the product, ethicon, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.A manufacturing record evaluation was performed for the finished device lot, and no non-conformances were identified.Attempts have been made to retrieve the device.To date the device has not been returned.If the device or further details are received at a later date a supplemental medwatch will be sent.Additional information was requested, and the following was obtained via: how many procedures/patients were affected by this issue? 4 procedures.Please provide the following information for each involved patient/procedure: how many devices demonstrated the alleged deficiency on each involved patient event? 4 stratafix were affected total (1 in each case/patient).When did the alleged deficiency occur (removal fr-om package / during handling prior to use on patient/ during passage through tissue / during tying / post-op)? if different, please explain.During passing through tissue.Were there any patient consequences? no.Please provide the status of the device(s) as it has not been received for analysis.If the device has been shipped, please provide the shipment tracking details.Shipping tracking#: (b)(4); confirmation#: (b)(4).Please provide the title of external person providing answers to follow-up (e.G.Nurse, surgeon, risk manager).Materials management.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
STRATAFIX
Common Device Name
SUTURE, SURGICAL, ABSORBABLE, POLYDIOXANONE
Manufacturer (Section D)
ETHICON INC.
1000 route 202
raritan NJ 08869
Manufacturer (Section G)
ETHICON INC.-SAN LORENZO PR
road 183, km 8.3
san lorenzo PR 00754
Manufacturer Contact
orla o'mahony
1000 route 202
raritan, NJ 08869
3035526892
MDR Report Key25164479
Report Number2210968-2026-05262
Device Sequence Number10191127
Product Code NEW
UDI-Device Identifier10705031464544
UDI-Public(01)10705031464544(17)270630(10)109E3D
Combination Product (Y/N)Y
Initial Reporter CountryUS
PMA/510(K) Number
K192144
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 06/16/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue NumberSXPP2B405
Device Lot Number109E3D
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 04/27/2026
Supplement Date Received by Manufacturer06/09/2026
Initial Report FDA Received Date05/13/2026
Supplement Report FDA Received Date06/16/2026
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured07/01/2025
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-