SENSORY MEDICAL INC. CUBBY BEDS; TECHNOLOGY HUB Back to Search Results
Catalog Number KIT10001
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 04/22/2026
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Sensory medical advised the parent to discontinue use of the bed until the issue can be resolved.Sensory medical provided a replacement cloth cover to the customer.Sensory medical provided a return shipping label for return of the damaged technology hub.It has not been received back for examination as of the writing of this investigation.Sensory medical has followed up on 04/23/2026, 04/24/2026, 04/27/2026, 04/28/2026, 05/04/2026, 05/06/2026 and 05/08/2026, to obtain additional information relevant to the investigation.Multiple statements are made in the cubby bed user manual and technology hub accessory manual regarding safe use of the bed to prevent elopement and other safety concerns.Page 3 of the user manual contains multiple warnings, including: "do not use this product if you do not understand or cannot follow the accompanying warnings and instructions." "do not allow anyone to climb or hang from the product." "you are responsible for providing adult supervision while using this product." "use the maintenance checklist in the manual to inspect the canopy, seams, zippers, electronic accessories, cords, metal frame, screws, slats, locks, and safety sheets every 30 days." "if any damage is present, immediately discontinue use and contact cubby for repair or replacement." "check this product for damaged hardware, loose joints, loose bolts or other fasteners, missing parts or sharp edges before and after assembly and frequently during use.Securely tighten loose bolts and other fasteners." "do not use product if any parts are missing, damaged or broken.Contact cubby for replacement parts and instructional literature if needed.Do not substitute parts." "never leave patient user unattended for extended periods of time without monitoring and sensory stimulation." "openings in and between bed parts can entrap the head and the neck." page 15 assembly + care faq's includes description and use of the locks on the safety sheets and the technology hub zippers and s-clips on the doors.On page 22 includes a monthly maintenance checklist, one check is: "each zipper on the canopy is functional and seams of zippers are intact" page 22 of the user manual, maintenance checklist, describes to discontinue use of the bed if anything is damaged or missing.Additional investigation is needed, and sensory medical's root cause investigation is ongoing.There was no report of injury as a result of the event.
 
Event or Problem Description
The parent reported that her son damaged the housing of the technology hub, causing it to crack and separate, and eloped from the bed through the portal area.The parent explained that she is unaware of how her child was able to break the technology hub housing.She confirmed that the padlock was installed on the technology hub at the time of the reported event.The parent provided one photo of the damaged technology hub.The photo confirms the housing is cracked, and a piece of the housing is broke off from the device.There was no report of injury as a result of the event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CUBBY BEDS
Common Device Name
TECHNOLOGY HUB
Manufacturer (Section D)
SENSORY MEDICAL INC.
1580 logan street
suite 500
denver CO 80203
Manufacturer (Section G)
SENSORY MEDICAL INC.
1580 logan street
suite 500
denver CO 80203
Manufacturer Contact
justin o'neal
1580 logan street
suite 500
denver, CO 80203
8559642664
MDR Report Key25164949
Report Number3016541541-2026-00038
Device Sequence Number16000216
Product Code KMI
UDI-Device Identifier00860007550263
UDI-Public860007550263
Combination Product (Y/N)N
Initial Reporter StateAZ
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Consumer
Initial Reporter Occupation Other
Type of Report Initial
Report Date (Section B) 05/13/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Catalogue NumberKIT10001
Device Lot Number12182023
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 04/23/2026
Initial Report FDA Received Date05/13/2026
Was Device Evaluated by Manufacturer? (Y/N) No
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Patient SexMale
-
-