BECTON DICKINSON & CO (FRANKLIN LAKES) BD VACUTAINER® LITHIUM HEPARINN (LH) 95 USP UNITS BLOOD COLLECTION TUBES; TUBES, VIALS, SYSTEMS, SERUM SEPARATORS, BLOOD COLLECTION
|
Back to Search Results |
|
| Catalog Number |
367886 |
| Medical Device Problem Code |
Short Fill (1575)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
04/17/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Investigation summary: bd did not receive any samples or photos for investigation.A total of 100 retained samples were visually inspected, and the hemogard closure assembly was found to be correctly assembled and placed on the tubes.Functional tests were performed on 10 retained samples.All tubes were within specification limits, and no issues were identified.Based on a review of the device history record for the incident lot, all product specifications and requirements for lot release were met.This complaint has not been confirmed for the indicated failure mode: underfill.Complaints received for this device and reported condition will continue to be tracked and trended.Information will be captured on trend reports and monitored.Our business team regularly reviews the collected data for the identification of emerging trends.
|
| |
|
Event or Problem Description
|
|
It was reported while using bd vacutainer® lithium heparinn (lh) 95 usp units blood collection tubes, the tube of one patient sample underfilled resulting in recollection.No adverse impact was reported regarding the patient or user.
|
| |
|