MAUDE Adverse Event Report: BOSTON SCIENTIFIC CORPORATION HURRICANE RX; CATHETER, BILIARY, DIAGNOSTIC

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC CORPORATION HURRICANE RX; CATHETER, BILIARY, DIAGNOSTIC Back to Search Results
Model Number M00545920
Medical Device Problem Codes Burst Container or Vessel (1074); Use of Device Problem (1670); Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 04/20/2026
Type of Reportable Event Malfunction
Event or Problem Description
Note: this report pertains to one of two hurricane dilatation balloon used in the same patient and procedure.It was reported to boston scientific corporation that two hurricane dilatation balloons were attempted to be used in the common bile duct during a dilatation procedure performed on (b)(6) 2026.During the procedure.It was reported that the balloon ruptured.Additionally, the same problem happened on the other hurricane dilatation balloon device.It was reported that no piece of the balloon detached inside the patient.The procedure was completed with another hurricane dilatation balloon.There were no patient complications reported as a result of this event.Note: the instructions for use (ifu) indicate that this balloon should not be pre-inflated prior to use in the procedure.However, the customer reported that the balloon was pre-inflated outside of the patient.
 
Additional Manufacturer Narrative
Block h6: imdrf device code a0402 captures the reportable event of a balloon burst.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HURRICANE RX
Common Device Name
CATHETER, BILIARY, DIAGNOSTIC
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
300 boston scientific way
marlborough MA 01752
Manufacturer (Section G)
BOSTON SCIENTIFIC CORK LIMITED
cork business technology park
model farm road
cork T12 Y K88
EI   T12 YK88
Manufacturer Contact
christina lam
300 boston scientific way
marlborough, MA 01752
9145716918
MDR Report Key25166481
Report Number3005099803-2026-01884
Device Sequence Number10567129
Product Code FGE
UDI-Device Identifier08714729283836
UDI-Public(01)08714729283836(17)270926(10)37559136
Combination Product (Y/N)N
Initial Reporter CountryTW
PMA/510(K) Number
K130484
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Other,Foreign,Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 06/16/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberM00545920
Device Catalogue NumberM00545920
Device Lot Number0037559136
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer05/07/2026
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 04/20/2026
Supplement Date Received by Manufacturer05/21/2026
Initial Report FDA Received Date05/13/2026
Supplement Report FDA Received Date06/16/2026
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured09/26/2025
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-