• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: FISHER & PAYKEL HEALTHCARE LTD MR850; RESPIRATORY HUMIDIFIER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

FISHER & PAYKEL HEALTHCARE LTD MR850; RESPIRATORY HUMIDIFIER Back to Search Results
Model Number MR850JHU
Medical Device Problem Code No Audible Alarm (1019)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(4).Method: the subject mr850 respiratory humidifier was received at our fisher & paykel (f&p) regional office in the usa where it was visually inspected and performance tested by a trained f&p healthcare service technician.Our investigation is based on the information provided by the f&p service technician, previous investigations of similar complaints, and our knowledge of the product.Results: performance testing of the subject mr850 respiratory humidifier confirmed that no sound was generated from the speaker.The unit was repaired and returned to the customer after passing functional and electrical safety tests.Conclusion: the root cause of the speaker fault was not determined.The mr850 respiratory humidifier product technical manual contains a maintenance schedule which instructs the user to conduct annual visual checking and performance testing of the mr850 respiratory humidifier.In addition, the product technical manual also states that "all servicing procedures should be followed by a humidifier test, and an electrical safety test to ensure proper operation".The mr850 respiratory humidifier is equipped with visual alarm indicators in addition to the audible alarm.
 
Event or Problem Description
During device evaluation of an mr850 respiratory humidifier at our fisher & paykel healthcare (f&p) regional office in the usa, it was found that the audible alarm was not functioning.There was no patient involvement reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MR850
Common Device Name
RESPIRATORY HUMIDIFIER
Manufacturer (Section D)
FISHER & PAYKEL HEALTHCARE LTD
15 maurice paykel place
east tamaki
auckland, 2013
NZ  2013
Manufacturer (Section G)
FISHER & PAYKEL HEALTHCARE LTD
15 maurice paykel place
east tamaki
auckland, 2013
NZ   2013
Manufacturer Contact
diego banuelos
17400 laguna canyon road
suite 300
irvine, CA 92618
9494341005
MDR Report Key25167900
Report Number9611451-2026-01638
Device Sequence Number10643729
Product Code BTT
UDI-Device Identifier09420012407290
UDI-Public(01)09420012407290(10)2101924418(11)211129
Combination Product (Y/N)N
Initial Reporter StateIL
Initial Reporter CountryUS
PMA/510(K) Number
K110019
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,Distributor
Initial Reporter Occupation Biomedical Engineer
Type of Report Initial
Report Date (Section B) 05/14/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberMR850JHU
Device Catalogue NumberMR850JHU
Device Lot Number2101924418
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 04/16/2026
Initial Report FDA Received Date05/14/2026
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured11/29/2021
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Patient SexUnknown
-
-