• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DATEX-OHMEDA, INC. AISYS; ANESTHESIA GAS MACHINE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DATEX-OHMEDA, INC. AISYS; ANESTHESIA GAS MACHINE Back to Search Results
Model Number CS2 WITH ET CONTROL
Medical Device Problem Code Mechanical Problem (1384)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 04/16/2026
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that there was a malfunction resulting in loss of mechanical ventilation during pre-use checkout.There was no patient involvement.
 
Additional Manufacturer Narrative
A ge healthcare service representative performed a checkout of the system and confirmed the reported issue.The sodasorb canister was replaced to resolve the issue.Block a: no report of patient involvement.Legal manufacturer: hcs madison - 3030 ohmeda dr, usa madison, wi 53718.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AISYS
Common Device Name
ANESTHESIA GAS MACHINE
Manufacturer (Section D)
DATEX-OHMEDA, INC.
3030 ohmeda dr,
madison, WI 53718
Manufacturer (Section G)
DATEX-OHMEDA, INC.
3030 ohmeda dr,
madison, WI 53718
Manufacturer Contact
michaela moore
3030 ohmeda drive
madison, WI 53718
MDR Report Key25172115
Report Number2112667-2026-03263
Device Sequence Number10474383
Product Code BSZ
UDI-Device Identifier00195278588128
UDI-Public(01)00195278588128(11)260106
Combination Product (Y/N)N
Initial Reporter StateCA
Initial Reporter CountryUS
PMA/510(K) Number
K170872
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,User Facility
Initial Reporter Occupation Biomedical Engineer
Type of Report Initial
Report Date (Section B) 05/14/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberCS2 WITH ET CONTROL
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 04/16/2026
Initial Report FDA Received Date05/14/2026
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured01/06/2026
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device A
Patient Sequence Number1
Patient SexUnknown
-
-