| Model Number |
FGS-0636 |
| Medical Device Problem Code |
Positioning Failure (1158)
|
| Health Effect - Clinical Code |
Airway Obstruction (1699)
|
| Date of Event |
04/20/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that the capsule failed to attach to the patient's esophagus after following the correct steps of 30 seconds suction of at least 550mmhg.Capsule was not at the desired location but not with the delivery device anymore when physician went in with scope to check.Deployed capsule found at the patient's mouth and was retrieved by hands.A second capsule of the same lot number was deployed successfully.There was no patient or user harm.
|
| |
|
Additional Manufacturer Narrative
|
|
H3 evaluation summary: medtronic conducted an investigation based upon all information received.The capsule and delivery system were returned for evaluation.Visual inspection of the delivery device and capsule found no notable conditions.It was reported that the bravo capsule failed to attach to the patient's esophagus.The reported issue was confirmed.The most likely cause could not be established from the information available.The manufacturing records for each device are thoroughly reviewed prior to release to ensure that it meets all medtronic quality specifications.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
| |
|
Search Alerts/Recalls
|