Model Number PCF-H190DI
Medical Device Problem Code Poor Quality Image (1408)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The device was returned to olympus for inspection and the reported failure was not confirmed.In addition, the following reportable malfunctions were identified during the device evaluation: light guide lens had foreign objects.Based on the results of the investigation, a root cause could not be identified.Should additional relevant information become available, a supplemental report will be submitted.Olympus will continue to monitor field performance for this device.
 
Event or Problem Description
It was reported that the colonovideoscope had blue vertical stripes on the screen.The issue was found during inspection for use.There were no reports of patient harm.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EVIS EXERA III COLONOVIDEOSCOPE
Common Device Name
COLONOVIDEOSCOPE
Manufacturer (Section D)
AIZU OLYMPUS CO., LTD.
3-1-1 niiderakita
aizuwakamatsu-shi, fukushima 965-8 520
JA  965-8520
Manufacturer (Section G)
AIZU OLYMPUS CO., LTD.
3-1-1 niiderakita
aizuwakamatsu-shi, fukushima 965-8 520
JA   965-8520
Manufacturer Contact
todd brill
800 west park drive
westborough, MA 01581
5082077661
MDR Report Key25174435
Report Number9610595-2026-39284
Device Sequence Number10199354
Product Code FDF
UDI-Device Identifier04953170363696
UDI-Public04953170363696
Combination Product (Y/N)N
Initial Reporter CountryNO
PMA/510(K) Number
N/A
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,User Facility
Initial Reporter Occupation Nurse
Type of Report Initial
Report Date (Section B) 05/14/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberPCF-H190DI
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/30/2026
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 04/28/2026
Initial Report FDA Received Date05/14/2026
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured10/01/2020
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Patient SexUnknown
-
-