| Catalog Number |
MCP427H |
| Medical Device Problem Code |
Adverse Event Without Identified Device or Use Problem (2993)
|
| Health Effect - Clinical Codes |
Erythema (1840); Hypersensitivity/Allergic reaction (1907); Unspecified Infection (1930); Local Reaction (2035)
|
| Date of Event |
01/01/2026
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
It was that a patient underwent a breast surgery on an unknown date and suture was used.Post-op, the patient experienced an infection around the suture a couple of days after the procedure.The surgeon opined this occurred after converting to triclosan-coated plus sutures and attributed the reactions/infections to the triclosan-coated plus suture.Redness and blotchiness were observed around the incision areas.Additional information is requested.
|
| |
|
Additional Manufacturer Narrative
|
|
Product complaint # (b)(4).This is a combination product, and the event has been reviewed for both the suture and the triclosan.This report is being submitted pursuant to the provisions of 21 cfr, part 803, part 4 subpart b.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by ethicon, or its employees that the report constitutes an admission that the product, ethicon, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.To date the device has not been returned.If the device or further details are received at a later date a supplemental medwatch will be sent.D4: full udi currently unavailable since the exact lot of the suspected device cannot be determined.Additional information: d4 - the actual device batch number associated with this event is not known.The international affiliate reports the following possible batch numbers: batch 10cph0 batch 10cj1z batch 10b292 batch 10c1ul batch 10csva attempts are being made to obtain the following information.To date no response has been provided.If further details are received at a later date a supplemental medwatch will be sent.Please provide the patient's demographic information including age, gender, weight, bmi at the time of index procedure.Date and name of index surgical procedure? the diagnosis and indication for the index surgical procedure? were any concomitant procedures performed? what was the initial approach for the index surgical procedure? (open, laparoscopic or other)? on what tissue was the suture used? what was the tissue condition (normal, thin, calcified, fragile, diseased)? for the original intended closure, how were all layers closed? how was the suture placed (interrupted or continuous)? how was the suture originally tied (multiple knots, square knot, etc.)? please describe the patient manifestations of the reported infection (location, severity, appearance).What symptoms did the patient experience following the index surgical procedure? onset date? please provide the onset date/time of infection from the initial surgical procedure.Were there any pre-existing signs/symptoms of active infection prior to this surgical procedure? did the patient receive any prophylactic antibiotics pre-, intra- or post-operation? were cultures and sensitivities performed? if so, please provide the results.How much and what type of drainage is present in this wound? please describe any medical intervention performed including medication name and results.Please describe the patient manifestations of the reported reaction (location, severity, appearance, generalized or local reaction).If generalized reaction, please provide each location, severity and appearance.Please provide the onset date/time of the reaction from the initial procedure.Does the patient have a known allergic history to any medical devices, food and/or medication? was allergy testing performed? if so, please describe with results.Did the patient undergo a patch test? are there any photos available? were any pre-op cleansing procedures changed recently? if yes, please describe did the operating surgeon observe any suture deficiency or anomaly before, during, after the suture placement or during any re-operation? other relevant patient history/concomitant medications? what is the physician¿s opinion as to the etiology of or contributing factors to this event? what is the patient's current status?.
|
| |
|
Additional Manufacturer Narrative
|
|
Product complaint: (b)(4).This is a combination product, and the event has been reviewed for both the suture and the triclosan.This report is being submitted pursuant to the provisions of 21 cfr, part 803, part 4 subpart b.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by ethicon, or its employees that the report constitutes an admission that the product, ethicon, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.Corrected information: h6 health effect - clinical code.Additional information was requested, and the following was obtained: i spoke to this account several times and they do not want to move forward with these product complaints due to the time involved, patient/procedure details needed and the lack of possible resolution since there were no issues uncovered from the lots used.A majority of the surgeons are currently running a ¿trial¿ and using non-plus suture on one breast and the triclosan suture on the other breast for all mastopexy¿s performed to try to rule out a triclosan allergy.They did share that these patients did not have an infection postoperatively, but more of an allergic reaction.
|
| |
|
Additional Manufacturer Narrative
|
|
Product complaint: (b)(4).This is a combination product, and the event has been reviewed for both the suture and the triclosan.This report is being submitted pursuant to the provisions of 21 cfr, part 803, part 4 subpart b.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by ethicon, or its employees that the report constitutes an admission that the product, ethicon, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.Additional information: h6.Additional information: d4, h4 - the actual device batch number associated with this event is not known.The international affiliate reports the following possible batch numbers: batch 10cph0 expiration date: nov/30/2027 date of mfg.: dec/15/2025.Batch 10cj1z expiration date: nov/30/2027 date of mfg.: dec/05/2025.Batch 10b292 expiration date: aug/31/2027 date of mfg.: sep/19/2025.Batch 10c1ul expiration date: oct/31/2027 date of mfg.: nov/05/2025.Batch 10csvaa invalid.A manufacturing record evaluation was performed for the possible finished device lots, and no non-conformances were identified.
|
| |
|