The reporter indicated that following an implantable collamer lens (icl) implantation procedure, the patient experienced excessive vault, elevated iop with pupil block and angle closure, significant reduction of iridoncoreanl angles, shallowing of anterior chamber, refractive surprise, refractive error and blurred vision.Topical drops administered for the elevated iop.This did not resolve the problem.The lens was explanted.On the same day in a separate surgery the lens was replaced with a same model, similar power but shorter length lens.The problem was resolved.Cause of the event was reported as device.There are multiple medical device reports associated with this patient.This report is 1 of 2 total reports.Based on the information provided, the risk assessment for our product, and our experience, we have determined the cause or contribution to the reported issue is not indicative of a manufacturing related issue or product deficiency.
|