Further information from the reporter regarding event, product, or patient details has been requested.No additional information is available at this time.A review of the device history record has been completed.No deviations or non-conformances noted.Reason for reoperation: rupture.
|
|
This report was impacted by emdr scheduled maintenance occurring july 22-27, 2026
device evaluation: the device related to the reported events rupture was received on june 15,2026, with lot number 1748563.Per the investigation procedure, the device is analyzed through visual microscopic inspection if openings are observed and a weight verification.Per the analysis performed, the assessments of the complaints and any potential manufacturing issue are displayed along with any further actions required:
- rupture: observed broken device through microscopic inspection assessed as fold flaw opening.No further actions are required as it is not related to the manufacturing process.None of the other observations performed during the device analysis (non-penetrating nick) are found to be potentially related to the manufacturing.Additional, changed, and/or corrected data: d.9., h.2., h.3., h.6.
|