| Model Number |
20000700700011010 |
| Medical Device Problem Codes |
Failure to Power Up (1476); Failure of Device to Self-Test (2937)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Zoll medical corporation has received the product and will be providing a supplemental report when our investigation is completed.
|
| |
|
Event or Problem Description
|
|
Complainant alleged that during functional testing, the device would intermittently not power up and prompted a "unit failed" message.Complainant indicated that there was no patient involvement in the reported malfunction.
|
| |
|
Additional Manufacturer Narrative
|
|
This supplemental medwatch report is reporting the evaluation of the device and correcting information submitted on the initial medwatch report.Please reference section d2.The device was returned to zoll medical corporation; the customer's report was duplicated and attributed to a defective transformer on the analog board.The analog board was replaced to resolve the report.The device was recertified and returned to the customer.Analysis for reports of this type has not identified an increase in trend.
|
| |
|
Search Alerts/Recalls
|