Additional Manufacturer Narrative
|
|
(b)(4).D10: cat# 00700006620 lot# 64985497 66mm cup size revision shell.Cat# 00713306658 lot# 3002378 provisional short flange cage.Cat# 00625006540 lot# j7629854 bone screw.Cat# 00625006515 lot# j7131464 bone screw.Cat# 00625006515 lot# j7583492 bone screw.Cat# 00625006520 lot# 63044691 bone screw.Cat# 00712306658 lot# 3089341 acetabular revision system - cage.Cat# 00625006550 lot# 64143387 bone screw.Cat# 00625006515 lot# j7583489 bone screw.Cat# 00625006540lot# j7622086 bone screw.Cat# 00625006520 lot# 65632422 bone screw.Cat# 00625006520 lot# j7576006 bone screw.Cat# 20123604 lot# 65051472 36mm i.D.Size d high wall liner.Cat# 650-0662 lot# 3144899 delta ceramic fem hd 36/+3mm.Cat# 110034355 lot# ax48ca0604 refobacin bc r 1x40 us.Cat# 00504905500 lot# 66090710 w/(2) allen medullary cement plugs.The device will not be returned for analysis; however, an investigation of the reported event is in progress.Once the investigation is completed, a supplemental medwatch 3500a will be submitted.
|
| |
|
Event or Problem Description
|
|
It was reported that approximately two months post-hip procedure, the patient presented with septic shock due to mrsa infection in the hip.They had fevers, difficulty walking, swelling, abscess, and positive cultures for mrsa and septic shock.The head was exchanged in an i&d procedure before the stage one revision one month later.Op records note purulent fluid, large fascial dehiscence.All components were replaced with a cement spacer during stage one without complications.The stage two revision took place four months later.Attempts have been made and no further information has been provided.
|
| |
|
Event or Problem Description
|
|
No further information is available at the time of this report.
|
| |
|
Additional Manufacturer Narrative
|
|
This follow-up report is being submitted to relay additional information.The following sections were updated: d4; h2; h4; h6; h10.The following sections were corrected: h6 component code
review of the reported event by a health care professional found:
during the investigation process a review of the sterile certifications were reviewed and found to be conforming with no applicable deviations.Devices were verified to have gone through acceptable sterilization process following iso/aami/astm & eu published guidelines.There are multiple factors that may contribute to an infection that are outside the control of zimmer biomet, such as external factors, i.E.Hospital/surgical environment, provider-related risk factors, and/or patient comorbidities/risk factors.As there are no indications of a product or process issues identified affecting implant safety or effectiveness, therefore implanted products are not identified as the source or contributing to the reported infection.Complaint is not confirmed.No product was returned or pictures provided; visual and dimensional evaluations could not be performed.Product will not be evaluated as it has been determined the event is not related to the device.The root cause of the reported event was determined to be unrelated to the device.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
|
| |
|