| Model Number |
04.01.0123 |
| Medical Device Problem Code |
Device Dislodged or Dislocated (2923)
|
| Health Effect - Clinical Code |
Joint Dislocation (2374)
|
| Date of Event |
05/01/2026
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Batch review performed on 05 may 2026 reverse shoulder system 04.01.0123 humeral reverse hc liner d 39/+3mm (k170452) lot: 2514686: (b)(4) items manufactured and released on 29-jul-2025.Expiration date: 10-jul-2030.No anomalies found related to the problem.To date, (b)(4) items of the same lot have been sold with no similar reported event during the period of review.Reverse shoulder system 04.01.0170 glenosphere - d 39x24.5 (k170452) lot: 2505967: (b)(4) items manufactured and released on 04-jun-2025.Expiration date: 22-may-2030.No anomalies found related to the problem.To date, (b)(4) items of the same lot have been sold with no similar reported event during the period of review.Root cause: dislocation is possible literature described adverse event after primary hip arthroplasties and causes are often unknown.There is no indication that any potential issue with the device may have caused or contributed to the event and the investigation does not indicate any potential manufacturing related issue or systemic deficiency.
|
| |
|
Event or Problem Description
|
|
At about 5 months from the primary, the patient came in presenting pain from a periprosthetic fracture at the base of the humeral stem due to a fall.The surgeon cabled the fracture and revised the metaphysis from+0mm/0° to +9mm/0° the short diaphysis to a long diaphysis, and the liner changed from 39/+0mmto d 39/+3mm.The surgery was completed successfully.About 2 months later, the patient presented with pain due to dislocation of the liner from the glenosphere and the cause is unknown.No signs of trauma were reported.The surgeon revised the hc liner from d 39/+3mm to a constrained d 39/+0mm.Surgery was completed successfully.
|
| |
|
Search Alerts/Recalls
|