• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDACTA INTERNATIONAL SA SHOULDER SYSTEM; HUMERAL REVERSE HC LINER Ø39/+3MM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDACTA INTERNATIONAL SA SHOULDER SYSTEM; HUMERAL REVERSE HC LINER Ø39/+3MM Back to Search Results
Model Number 04.01.0123
Medical Device Problem Code Device Dislodged or Dislocated (2923)
Health Effect - Clinical Code Joint Dislocation (2374)
Date of Event 05/01/2026
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Batch review performed on 05 may 2026 reverse shoulder system 04.01.0123 humeral reverse hc liner d 39/+3mm (k170452) lot: 2514686: (b)(4) items manufactured and released on 29-jul-2025.Expiration date: 10-jul-2030.No anomalies found related to the problem.To date, (b)(4) items of the same lot have been sold with no similar reported event during the period of review.Reverse shoulder system 04.01.0170 glenosphere - d 39x24.5 (k170452) lot: 2505967: (b)(4) items manufactured and released on 04-jun-2025.Expiration date: 22-may-2030.No anomalies found related to the problem.To date, (b)(4) items of the same lot have been sold with no similar reported event during the period of review.Root cause: dislocation is possible literature described adverse event after primary hip arthroplasties and causes are often unknown.There is no indication that any potential issue with the device may have caused or contributed to the event and the investigation does not indicate any potential manufacturing related issue or systemic deficiency.
 
Event or Problem Description
At about 5 months from the primary, the patient came in presenting pain from a periprosthetic fracture at the base of the humeral stem due to a fall.The surgeon cabled the fracture and revised the metaphysis from+0mm/0° to +9mm/0° the short diaphysis to a long diaphysis, and the liner changed from 39/+0mmto d 39/+3mm.The surgery was completed successfully.About 2 months later, the patient presented with pain due to dislocation of the liner from the glenosphere and the cause is unknown.No signs of trauma were reported.The surgeon revised the hc liner from d 39/+3mm to a constrained d 39/+0mm.Surgery was completed successfully.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SHOULDER SYSTEM
Common Device Name
HUMERAL REVERSE HC LINER Ø39/+3MM
Manufacturer (Section D)
MEDACTA INTERNATIONAL SA
strada regina
castel san pietro 6874
SZ  6874
Manufacturer Contact
marco giannessi
strada regina
castel san pietro 6874
SZ   6874
MDR Report Key25185355
Report Number3005180920-2026-00466
Device Sequence Number16125457
Product Code PHX
UDI-Device Identifier07630040706308
UDI-Public07630040706308
Combination Product (Y/N)N
PMA/510(K) Number
K170452
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2026
Device Explanted Year2026
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Company Representative
Initial Reporter Occupation Other
Type of Report Initial
Report Date (Section B) 05/15/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Model Number04.01.0123
Device Lot Number2514686
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 05/01/2026
Initial Report FDA Received Date05/15/2026
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured07/29/2025
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient SexMale
Patient RaceWhite
-
-