Model Number FNAV-DS-LG
Medical Device Problem Codes Retraction Problem (1536); Material Deformation (2976)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 04/21/2026
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The device is not returning for evaluation.Investigation is not yet complete.A follow-up report will be submitted with all additional relevant information.
 
Event or Problem Description
Clinical information: crd_1059 - vista nova study, patient site id: site id: (b)(6) (ae-r995117901, dm-r995171201, dm-r995171001).It was reported that on (b)(6) 2026, a 29mm navitor vision valve was chosen for implant using a large flexnav delivery system.The activated clotting levels were 220,294s and heparin was given.There was not enough calcification.A pre-implant balloon valvuloplasty was done with a 23mm non-abbott balloon.After 1st deployment attempt the physician decided to re-sheath the valve because was too high.During the re-sheath the capsule was deformed and was impossible to re-sheath completely the valve.The valve and delivery system were removed and a new 29mm navitor vision valve and large flexnav delivery system were chosen and successfully implanted.The final implant depth at non-coronary cusp (ncc) was 6.08mm and at left coronary cusp (ncc) was 8.25mm.A post-implant balloon valvuloplasty was performed due to high mean gradient.After the valve implant, the patient had a complete heart block and permanent pacemaker was implanted.The patient was reported discharged.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FLEXNAV DELIVERY SYSTEM
Common Device Name
AORTIC VALVE, PROSTHESIS, PERCUTANEOUSLY DELIVERED
Manufacturer (Section D)
ABBOTT MEDICAL
5050 nathan lane n
plymouth MN 55442
Manufacturer (Section G)
ABBOTT MEDICAL REG#3007113487
177 county road b east
st. paul MN 55117 1789
Manufacturer Contact
karen krouse
5050 nathan lane n
plymouth, MN 55442
6517565400
MDR Report Key25185548
Report Number2135147-2026-03254
Device Sequence Number10755699
Product Code NPT
UDI-Device Identifier05415067031389
UDI-Public(01)05415067031389(17)271231(10)11205719
Combination Product (Y/N)N
Initial Reporter CountryIT
PMA/510(K) Number
P190023
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Study,Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 06/29/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberFNAV-DS-LG
Device Catalogue NumberFNAV-DS-LG
Device Lot Number11205719
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 04/22/2026
Supplement Date Received by Manufacturer06/18/2026
Initial Report FDA Received Date05/15/2026
Supplement Report FDA Received Date06/29/2026
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured01/10/2026
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient Age76 YR
Patient SexMale
Patient Weight68 KG
-
-