| Model Number |
DDU-A2300 |
| Medical Device Problem Code |
Failure to Power Up (1476)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Analysis of the aed's log files identified that the customer's failure to promptly address red asi warnings reduced battery life expectancy.On 11/01/20 the aed identified that the aed pads needed to be replaced and attempted to alert the user by flashing its red asi and chirping.The aed continued to flash and chirp until 3/16/21 when replacement pads were installed.On 11/24/23 the aed identified that the aed battery was getting low and attempted to alert the user by flashing its red asi and chirping.The aed continued to flash and chirp until 12/07/23 when the battery pack became completely depleted.There was no evidence in the electronic logs of any human interaction to address the aeds condition until 1/23/24 when a replacement battery pack was inserted into the aed.
|
| |
|
Event or Problem Description
|
|
A customer reported that upon inserting a new battery pack, their aed powered on and prompted a service code.The customer did not report this occurring during patient use.
|
| |
|