It was reported that a patient underwent a cardiac ablation procedure with a thermocool smarttouch sf and during the operation, force value jumps from 4 grams to 50 after starting the ablation.The force curve was exactly the same, like the power curve on smartablate generator.Cable replacement did not fix the issue.When the catheter was replaced, the issue was resolved.There was no adverse event reported on the patient.This event was originally assessed as not mdr reportable.The device was returned to johnson & johnson medtech (j&j medtech) for evaluation.Following j&j medtech procedures, a visual inspection and force test of the returned device were performed.Visual inspection revealed reddish material and a hole in the surface of the pebax.This finding is mdr reportable.
|
|
The device was returned to johnson & johnson medtech (j&j medtech) for evaluation.Following j&j medtech procedures, a visual inspection and force test of the returned device were performed.Visual inspection revealed reddish material and a hole in the surface of the pebax.The device was connected to the carto 3 system and no errors were observed.The force feature of the device was evaluated, and the force values and the vector were observed within specifications.No force issues were observed.A manufacturing record evaluation was performed for the finished device batch number, and no internal actions were identified.The reddish material inside the pebax could be related to the force issue reported by the customer; therefore, the customer complaint was confirmed.The potential cause of the pebax damage could be related to the manipulation of the device during the procedure; however, this cannot be conclusively determined.The instructions for use (ifu) contain the following recommendations: do not scrub or twist the tip electrode because damage to the tip electrode bond may occur and loosen the tip electrode, or damage the contact force sensor and affect measurement accuracy.In addition, in order to prevent damage to the catheter tip, verify compatibility between the sheath and catheter, advance the catheter through the sheath prior to insertion.Any sheath < 8.5 f is contraindicated.As part of johnson & johnson medtech¿s quality process, all devices are manufactured, inspected, and released to approved specifications.If additional information is received regarding this event, a supplemental 3500a report will be submitted to the fda.Manufacturer's reference number: (b)(4).
|