| Model Number |
1948/52 |
| Medical Device Problem Code |
No Apparent Adverse Event (3189)
|
| Health Effect - Clinical Code |
Insufficient Information (4580)
|
| Date of Event |
05/12/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
No complaint was received with the return of the device.Failure event observed during analysis.Final analysis found as received, only the connector portion of the lead was returned in one piece.Visual examination of the lead found full breached degradation breaching the outer insulation and exposing the outer coil located adjacent to the connector boot.Electrical testing of the lead did not find any indication of conductor fractures or internal shorts.
|
| |
|
Event or Problem Description
|
|
This report is to advise of an event observed during analysis.
|
| |
|