| Model Number |
4968-35 |
| Medical Device Problem Codes |
Over-Sensing (1438); High Capture Threshold (3266)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
04/21/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
D10 continued: dtpb2d1 icd implanted: (b)(6) 2022.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
| |
|
Event or Problem Description
|
|
It was reported that the epicardial right ventricular (rv) lead triggered a lead integrity alert (lia) for high rate-non-sustained e pisodes and sensing integrity counter (sic) in addition to high threshold and t-wave oversensing (twos) resulting in inhibition of pacing; post device defibrillation.It was further reported that the epicardial rv defibrillator lead exhibited out of range (oor) high impedance.The leads remain in use.No patient complications have been reported as a result of this event.
|
| |
|