• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BOSTON SCIENTIFIC CORPORATION CAPIO SLIM; SPECIALIZED SURGICAL INSTRUMENTATION FOR USE WITH UROGYNECOLOGIC SURGICAL MESH

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC CORPORATION CAPIO SLIM; SPECIALIZED SURGICAL INSTRUMENTATION FOR USE WITH UROGYNECOLOGIC SURGICAL MESH Back to Search Results
Model Number M0068318261
Medical Device Problem Code Misfire (2532)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 01/12/2026
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Block h6: imdrf impact code f1901 captures the reportable event of additional surgical intervention.Imdrf device code a050502 captures the reportable event of device misfired.
 
Event or Problem Description
It was reported that during sacrospinous ligament fixation procedure, five capio slim devices were used for pelvic organ prolapse.During the procedure, the devices did not fire properly.Hospital admission was required for the patient for additional surgical intervention.No patient complications were reported.Note: this manufacturer report pertains to the second of five devices used during the procedure.
 
Event or Problem Description
It was reported that during sacrospinous ligament fixation procedure, five capio slim devices were used for pelvic organ prolapse.During the procedure, the devices did not fire properly.Hospital admission was required for the patient for additional surgical intervention.No patient complications were reported.Note: this manufacturer report pertains to the second of five devices used during the procedure.
 
Additional Manufacturer Narrative
Block h6: imdrf impact code f1901 captures the reportable event of additional surgical intervention.Imdrf device code a050502 captures the reportable event of device misfired.Block h11: the complaint device was not returned; therefore, no physical or visual analysis could be performed.As a result, the reported device performance allegation could not be confirmed.The device history record (dhr) confirmed that the device met all material, assembly and performance specification.A risk review was completed and confirmed that the event of "device misfire" was defined in the risk documentation.This event type has been accounted for during product risk analysis to support acceptable risk benefit for the product.Based on the information provided, the most probable cause of the reported issue could not be determined due to a lack of evidence, and in the absence of device evaluation or objective evidence, the most probable causes that may have contributed to the event remain unknown; therefore, this complaint is assigned an investigation conclusion code of "cause not established", as the investigation findings do not clearly confirm the presence or absence of the reported problem with the device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CAPIO SLIM
Common Device Name
SPECIALIZED SURGICAL INSTRUMENTATION FOR USE WITH UROGYNECOLOGIC SURGICAL MESH
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
300 boston scientific way
marlborough MA 01752
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
2546 calle primera
alajuela
CS  
Manufacturer Contact
farshad fahimi
4100 hamline avenue north
building c
saint paul, MN 55112
MDR Report Key25194154
Report Number2124215-2026-26532
Device Sequence Number8889829
Product Code PWI
UDI-Device Identifier08714729842224
UDI-Public(01)08714729842224(10)34886311
Combination Product (Y/N)N
Initial Reporter StateIL
Initial Reporter CountryUS
PMA/510(K) Number
K172060
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,User Facility,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 06/24/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberM0068318261
Device Catalogue Number831-826
Device Lot Number0034886311
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 04/17/2026
Supplement Date Received by Manufacturer05/25/2026
Initial Report FDA Received Date05/15/2026
Supplement Report FDA Received Date06/24/2026
Date Device Manufactured09/19/2024
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age74 YR
Patient SexFemale
Patient Weight162 LB
Patient RaceWhite
-
-