| Model Number |
P1840D |
| Medical Device Problem Code |
Positioning Failure (1158)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
04/20/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
During servicing by baxter, technical service identified that totalcare bariatric capital head section of bed was not going up.This occurred during baxter servicing/testing.There was no patient/user injury, medical intervention, symptom, or adverse event reported.There were no other devices reported to be involved.
|
| |
|
Additional Manufacturer Narrative
|
|
The baxter technician found the solenoid valve needed to be replaced.Per the baxter service manual the totalcare bed system requires an effective maintenance program.We recommend that you perform a semi-annual preventative maintenance.Preventative maintenance will minimize downtime due to excessive wear.Test the head section slide pivots, self-lubricating bearing, and slider pivot extrusions.Look for damage, and make sure the head section operates correctly.Replace components as necessary.A search of the baxter maintenance records showed baxter did not performed any preventative maintenance on thisbed.It is unknown if the facility performed any other preventative maintenance on this bed.The technician replaced the solenoid valve to resolve the reported event.Based on this information, no further action is required.
|
| |
|