|
Intuitive surgical, inc.(isi) did receive the single port monopolar curved scissors instrument to perform failure analysis.The instrument was found to have a cracked tube adapter.This component is located at the distal end of the instrument and serves as the interface between the instrument and the user installed tip accessory.The cracked piece, measuring approximately 2.9mm in length.The cracked tube adapter is likely the cause of the broken electrical contacts.Visual inspection was performed and there was no damage to the snake wrist cables, housing and shaft.Each input disk was manually articulated and responded accordingly.The release buttons were functional.The probable root cause is attributed to excessive force applied to an installed tip accessory during use, such as inadvertent collisions with other instruments.Additional common causes include improper installation or removal of the tip accessory.Additional observation(s) related to the customer reported complaint: the instrument was found to have broken contacts.The electrical contacts were located at the distal tip of the instrument and was observed broken inside the tube adapter.The electrical contacts measure approximately 2mm x 2.7mm and was not returned with the instrument.The broken contacts are likely cause by the cracked tube adapter.The root cause is attributed to use conditions.The instrument was found to have a detached fragment.During visual inspection, the detached fragment was observed from the broken electrical contacts.The area of detachment measures approximately 2mm x 2.7mm and was not returned with the instrument.Additional observation(s) not reported by site: the instrument was found to have residue.Visual inspection was preformed and white residue was observed on two of the clamping pulleys, located inside the backend of the instrument.The probable root cause is attributed to improper cleaning or sterilization techniques used during reprocessing, which may involve prolonged exposure of the instrument to harsh cleaning agents or insufficient flushing.The instrument was found to have multiple corroded bearings.The housing was removed for inspection, and orange discoloration was observed on eight of the bearings near the clamping pulley, which indicates corrosion.The root cause is attributed to use conditions.
|