| Model Number |
72404256 |
| Medical Device Problem Code |
Migration (4003)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
01/13/2026
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Model number/catalog number: 720185-01.Serial number: n/a.Batch/lot number: 1100201972.Model/catalog description: reservoir flat iz 100 ml.Unique identifier (udi) #: (b)(4).The device is not available for analysis; therefore, no physical or visual analysis of the product could be performed.The reported migration is a known risk associated with this device type and indicated as such in the instructions for use.
|
| |
|
Event or Problem Description
|
|
It was reported that the patient with this inflatable penile prosthesis (ipp) experienced herniation of the right cylinder and was dissatisfied with the device.A surgical procedure was performed in which the existing device was explanted and replaced.There were no further patient complications reported.
|
| |
|