• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MED-EL ELEKTROMEDIZINISCHE GERAETE GMBH SONATA; Cochlear Implant

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MED-EL ELEKTROMEDIZINISCHE GERAETE GMBH SONATA; Cochlear Implant Back to Search Results
Model Number SONATA
Medical Device Problem Code Circuit Failure (1089)
Health Effect - Clinical Code Head Injury (1879)
Date of Event 05/13/2026
Type of Reportable Event Malfunction
Event or Problem Description
The hearing performance with the device is affected.
 
Additional Manufacturer Narrative
The device has not been explanted.If it should be explanted, it should be returned to the manufacturer for evaluation.When available, a device failure analysis will be submitted as a follow up report.
 
Additional Manufacturer Narrative
Conclusion: the device investigation revealed mechanical damage to the stimulator electronics that is typical for severe external impact to the housing.The problems described in the recipient report and the reported accident appears to match the damage found.This is a final report.
 
Event or Problem Description
Following a trauma to the head a few weeks prior, the user no longer had hearing benefit with the device.The user has been re-implanted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SONATA
Common Device Name
Cochlear Implant
Manufacturer (Section D)
MED-EL ELEKTROMEDIZINISCHE GERAETE GMBH
innsbruck
AU 
Manufacturer Contact
laura simonotti
fuerstenweg 77a
innsbruck 6020
AU   6020
+43()577885626
MDR Report Key25197778
Report Number9710014-2026-00450
Device Sequence Number18662744
Product Code MCM
UDI-Device Identifier09008737049119
Combination Product (Y/N)N
Initial Reporter CountryTU
PMA/510(K) Number
P000025
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2017
Device Explanted Year2026
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Company Representative
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup,Followup
Report Date (Section B) 09/09/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Expiration Date04/04/2020
Device Model NumberSONATA
Was Device Available for Evaluation? Yes
Date Returned to Manufacturer08/04/2026
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 05/13/2026
Supplement Date Received by Manufacturer05/13/2026
05/13/2026
Initial Report FDA Received Date05/18/2026
Supplement Report FDA Received Date06/18/2026
09/09/2026
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured04/04/2017
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age10 YR
Patient SexMale
-
-