| Catalog Number |
SAGL2043 |
| Medical Device Problem Codes |
Positioning Failure (1158); Naturally Worn (2988)
|
| Health Effect - Clinical Codes |
Erosion (1750); Metal Related Pathology (4530)
|
| Date of Event |
04/23/2026
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
It was reported that the patient underwent an initial implantation approximately four (4) years and six (6) months ago.Subsequently, it was noted post-operatively that the tm modular peg of the glenoid penetrated the poly liner and protruded.The patient underwent a revision approximately four (4) years and five (5) months post-implantation where the initial plan was to convert from a tsa to rtsa; however, due to metallosis and severe bone resorption, implanting a glenoid implant was unable to occur.Therefore, bone was harvested from the ilium and used for a bone graft at the site of the resorption.It is noted the patient was very active, in which this event has caused post-op management difficult.Attempts have been made and no further information has been provided.
|
| |
|
Additional Manufacturer Narrative
|
|
(b)(4) d10: item# sagp0002; lot# 64783943 , item# 118001; lot# j7061058 , item# 113134; lot# 524680.G2: foreign - event occurred in japan.Customer has indicated that the product is in process of being returned to zimmer biomet for investigation.Once the investigation has been completed, a follow-up mdr will be submitted.If any further information which would change or alter any conclusions or information, a supplemental report will be filed accordingly.Zimmer biomet will continue to monitor for trends.
|
| |
|