| Catalog Number |
198714111 |
| Medical Device Problem Code |
Migration (4003)
|
| Health Effect - Clinical Codes |
Pain (1994); Loss of Range of Motion (2032)
|
| Date of Event |
04/30/2026
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
It was reported that patient presented in the er with poor function and pain.Femoral component appeared to have rotated roughly 90 degrees internally.The stem, sleeve and lps adapter were placed initially and left in place when converted from a spacer to a rev tka using an ips distal femur on (b)(6) 2026.There appeared to be damage on the taper of the adapter and unfortunately it could not be removed.The stem, sleeve and adapter were all removed and new components were placed with no delay nor patient harm.Doil: on (b)(6) 2025, dor: on (b)(6) 2026, affected side: right knee.
|
| |
|
Additional Manufacturer Narrative
|
|
Product complaint#: (b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803 (and/or part 4, as applicable).This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes, or its employees that the report constitutes an admission that the product, depuy synthes, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
|
| |
|
Additional Manufacturer Narrative
|
|
Product complaint # (b)(4).Investigation summary: no device associated with this report was received for examination.The product investigation found no evidence suspecting an error in the manufacturing or material that would be a contributing factor in the reported allegation(s).A records evaluation (mre) was not perform.As part of our company quality system process, all devices are manufactured, inspected, and distributed to approved specifications.Additional complaint information monitoring for potential safety signals will be conducted through complaint trending as part of the post-market surveillance.If additional information is made available, the investigation will be updated as applicable.Device history lot: there is no evidence suspecting an error in the manufacturing or material that would be a contributing factor in the reported allegation(s).A manufacturing records evaluation (mre) was not performed.Added: d1, d10 concomitant.
|
| |
|
Search Alerts/Recalls
|