MEDTRONIC PUERTO RICO VILLALBA ATTAIN PERFORMA MRI SURESCAN; DRUG ELUTING PERMANENT LEFT VENTRICULAR (LV) P Back to Search Results
Model Number 429878
Medical Device Problem Code Failure to Capture (1081)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 04/24/2026
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that the cardiac resynchronization therapy defibrillator (crt-d) inappropriately withheld therapy for a ventricular arrhythmia.It was further reported that the right atrial (ra) lead exhibited undersensing resulting in unnecessary pacing at times, the left ventricular (lv) lead exhibited a loss of capture post therapy delivery, and the right ventricular (rv) lead exhibited undersensing resulting in inappropriate pacing.The crt-d, ra lead, lv lead and rv lead remain in use.No patient complications have been reported as a result of this event.
 
Additional Manufacturer Narrative
Continuation of d10: product id: dtpa2qq (rtc626633s); product type: 0236-crt-d; implant date: on (b)(6) 2023; product id: 6935m62 (tdl543944v); product type: 0264-lead-hv; implant date: on (b)(6) 2023; product id: 5076-52 (pjn8766201); product type: 0270-lead-lv-pace; implant date: on (b)(6) 2023.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ATTAIN PERFORMA MRI SURESCAN
Common Device Name
DRUG ELUTING PERMANENT LEFT VENTRICULAR (LV) P
Manufacturer (Section D)
MEDTRONIC PUERTO RICO VILLALBA
road 149 km 56.3
villalba PR 00766
Manufacturer (Section G)
MEDTRONIC PUERTO RICO VILLALBA
road 149 km 56.3
villalba PR 00766
Manufacturer Contact
gerard torenvliet
8200 coral sea st ne
mounds view, MN 55112-4391
7635269277
MDR Report Key25209760
Report Number2649622-2026-14450
Device Sequence Number10536874
Product Code OJX
UDI-Device Identifier00643169601826
UDI-Public(01)00643169601826(17)240325
Combination Product (Y/N)N
Initial Reporter StateOH
Initial Reporter CountryUS
PMA/510(K) Number
P080006
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2023
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 05/18/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Expiration Date03/25/2024
Device Model Number429878
Device Catalogue Number429878
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 04/24/2026
Initial Report FDA Received Date05/18/2026
Date Device Manufactured04/18/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
DTPA2QQ CRTD.
Patient Age74 YR
Patient SexUnknown
Patient Weight34 KG
-
-