Catalog Number CEB231010A
Medical Device Problem Code Patient-Device Incompatibility (2682)
Health Effect - Clinical Codes Aneurysm expansion (4597); Clinical Endoleak (4843)
Date of Event 07/18/2023
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
H6: a review of the manufacturing records for the device verified that the lot met all pre-release specifications.W.L.Gore & associates, inc.(gore) is submitting this report to comply with 21 c.F.R.Part 803, the medical device reporting regulation.This report is based upon information obtained by gore, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Blank fields present on this report include required fields and fields determined to be not applicable.Blank required fields indicate that the information was not provided, was deemed unavailable or was not applicable.This report does not constitute an admission or a conclusion by fda, gore, or its associates that the device, gore or its associates caused or contributed to the event described in the report.In particular, this report does not constitute a legal admission by anyone that the product described in this report has any defects or has malfunctioned, as defined from a legal standpoint.These words are included in the report and are fixed items for selection created by the fda, to categorize the type of event solely for the purpose of reporting pursuant to part 803.This statement should be included with any information or report disclosed to the public under the freedom of information act.
 
Event or Problem Description
The following information was reported to gore: on (b)(6) 2023, the patient underwent endovascular treatment for an abdominal aortic aneurysm using gore® excluder® conformable aaa endoprosthesis and gore® excluder®aaa endoprosthesis devices.Gore® excluder® iliac branch endoprosthesis (ibe) was placed on the left side.Completion angiography following stent graft deployment revealed a type ii endoleak, but no additional intervention was performed, and the procedure was completed.The aneurysm diameter was approximately 67 mm.On an unknown date, postoperative follow-up imaging demonstrated enlargement of the aneurysm sac.On april 2026 (exact date unknown), computed tomography (ct) showed that the aneurysm diameter had increased to 77 mm.Delayed-phase ct imaging confirmed a type ii endoleak originating from a lumbar artery.The findings suggested aneurysm sac enlargement associated with the type ii endoleak.However, a linear contrast enhancement extending from the right leg of the stent graft was also observed, and the possibility of a type iiib endoleak could not be ruled out.On (b)(6) 2026, reintervention was performed.Intraoperative angiography identified a type ii endoleak from a lumbar artery.Embolization was planned for a later date due to difficulty accessing the responsible vessel.An additional stent graft was deployed within the right leg for preventive treatment, and the procedure was completed at that time.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
GORE® EXCLUDER® ILIAC BRANCH ENDOPROSTHESIS
Common Device Name
SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT
Manufacturer (Section D)
W. L. GORE & ASSOCIATES, INC.
1505 n. fourth street
flagstaff AZ 86004
Manufacturer (Section G)
MEDICAL PHOENIX 2 B/P
32470 n. north valley parkway
phoenix AZ 85085
Manufacturer Contact
rie kaba
1505 n. fourth street
flagstaff, AZ 86004
9285263030
MDR Report Key25212099
Report Number3007284313-2026-04789
Device Sequence Number18740761
Product Code MIH
UDI-Device Identifier00733132635283
UDI-Public(01)00733132635283(17)251013
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
P020004
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2023
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 05/19/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date10/13/2025
Device Catalogue NumberCEB231010A
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 05/01/2026
Initial Report FDA Received Date05/19/2026
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured10/15/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age88 YR
Patient SexFemale
-
-