| Model Number |
MR8-AVS14 |
| Medical Device Problem Code |
Difficult to Insert (1316)
|
| Health Effect - Clinical Codes |
Insufficient Information (4580); No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
04/10/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that the device had a difficulty in insertion of tool.No known patient impact reported.Additional information received that during evaluation the bearings were found to be misaligned.
|
| |
|
Additional Manufacturer Narrative
|
|
H3: product analysis : evaluation determined that difficult to insert the tool is been confirmed.The likely cause of failure is due to bearings misaligned.It was also noted that the bearings were corroded and the laser markings were faded.B.3.Date of incident or estimated date of incident was not provided to the manufacturer.Event notified date used as date of incident until further information is obtained.G.3.Aware date used as date of analysis performed date as the event is reported based on evaluation findings.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
| |
|