| Model Number |
MR8-AS10-R |
| Medical Device Problem Code |
No Apparent Adverse Event (3189)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
01/28/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
H3: product analysis:evaluation determined that bearing fallout at tube distal and that causes rattling sound.The likely cause of failure couldn't able to determine.It was also noted that tube and base markings are vague and need etching.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
| |
|
Event or Problem Description
|
|
Medtronic received information regarding a repair request and no alleged product deficiencies were reported.There was no patient involvement reported.Additional information was received stating that bearing fallout at tube distal that causes rattling sound were found during evaluation.
|
| |
|
Search Alerts/Recalls
|