• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: THOMMEN MEDICAL AG NEVO RC INICELL, PF 4.0, E4.0, L11.0MM; DENTAL IMPLANT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

THOMMEN MEDICAL AG NEVO RC INICELL, PF 4.0, E4.0, L11.0MM; DENTAL IMPLANT Back to Search Results
Model Number 4.23.024
Medical Device Problem Code Failure to Osseointegrate (1863)
Health Effect - Clinical Code Inadequate Osseointegration (2646)
Date of Event 01/16/2026
Type of Reportable Event Serious Injury
Event or Problem Description
Primary stability not achieved, implant wasn't completely covered with bone, no thread tap used, complication in site preparation, inadequate bone quality/quantity, mobility - another size of implant was chosen.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
NEVO RC INICELL, PF 4.0, E4.0, L11.0MM
Common Device Name
DENTAL IMPLANT
Manufacturer (Section D)
THOMMEN MEDICAL AG
neckarsulmstrasse 28
grenchen 2540
CH  2540
Manufacturer (Section G)
THOMMEN MEDICAL AG
neckarsulmstrasse 28
grenchen 2540
CH   2540
Manufacturer Contact
astrid weber
neckarsulmstrasse 28
grenchen 2540
CH   2540
MDR Report Key25214777
Report Number3003184527-2026-00950
Device Sequence Number15801781
Product Code DZE
UDI-Device Identifier07640182643430
UDI-Public07640182643430
Combination Product (Y/N)N
Initial Reporter CountryCH
PMA/510(K) Number
K093615
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2026
Device Explanted Year2026
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Dentist
Type of Report Initial
Report Date (Section B) 04/29/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Model Number4.23.024
Device Lot Number34603
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 04/29/2026
Initial Report FDA Received Date05/19/2026
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured05/16/2024
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient Age78 YR
Patient SexFemale
Patient EthnicityNon Hispanic
Patient RaceWhite
-
-