|
There was an allegation of questionable elecsys toxo igm assay results for two patient samples tested on a cobas e 411 analyzer (rack system).Sample 1: the initial results were consistently positive.The sample was sent to the external service lab and tested on a liaison system, which yielded negative results.Sample 2: on (b)(6) 2026, the initial results were consistently positive.The sample was sent to the external service lab and tested on a liaison system, which yielded negative results.During the (b)(6) 2026 test, it was confirmed that the quality control was not performed.
|