| Catalog Number |
496S085 |
| Medical Device Problem Codes |
Detachment of Device or Device Component (2907); Adverse Event Without Identified Device or Use Problem (2993)
|
| Health Effect - Clinical Code |
Insufficient Information (4580)
|
| Date of Event |
04/21/2026
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Additional information has been requested and if received, will be provided in the supplemental report.
|
| |
|
Event or Problem Description
|
|
As reported: "a case in which evolve was implanted on april 7.Postoperative follow-up x-rays confirmed separation between the stem and the head.Revision surgery is scheduled for april 28.".
|
| |
|
Event or Problem Description
|
|
As reported: "a case in which evolve was implanted on(b)(6).Postoperative follow-up x-rays confirmed separation between the stem and the head.Revision surgery is scheduled for (b)(6).".
|
| |
|
Additional Manufacturer Narrative
|
|
Correction - please refer to h6 device code.The reported event could not be confirmed based on evaluation available x-rays.The device inspection was not possible as the product was not returned for investigation.A review of the device history for the reported lot did not indicate any abnormalities.No corrective actions are required currently.No indications of material, manufacturing or design-related problems were found during the investigation.A review of the labeling did not indicate any abnormalities.Upon review of available x-rays medical expert opined that
the dissociation of the radial head from the stem can be confirmed based on the available radiographic images.Review of the x-rays does not indicate an apparent device-related failure.However, a definitive determination would require examination of the explanted device itself.One possible explanation for the observed dissociation is malalignment of the radial head implant center relative to the articulating surface of the radiocapitellar joint.Such mispositioning may generate eccentric joint loading and levering forces across the implant construct, potentially resulting in separation of the radial head from the stem over time.More detailed information about the complaint event as well as the affected device must be available to determine the root cause of the complaint event.If the device is returned or if any additional information is provided, the investigation will be reassessed.
|
| |
|