• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: NEURAVI LTD. EMBOGUARD; CATHETER, PERCUTANEOUS, NEUROVASCULATURE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

NEURAVI LTD. EMBOGUARD; CATHETER, PERCUTANEOUS, NEUROVASCULATURE Back to Search Results
Catalog Number BG8795U
Medical Device Problem Code Material Twisted/Bent (2981)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 05/02/2026
Type of Reportable Event Malfunction
Event or Problem Description
A healthcare professional reported that during a mechanical thrombectomy to treat acute ischemic stroke (ais), a 0.035 guidewire, an inner catheter, and an emboguard 87, 95 cm (product code: bg8795u, lot number: 10293947) were advanced to the common carotid artery.Upon removal of the inner catheter, a kink was observed on the emboguard, so it was replaced with a new balloon guiding catheter.The procedure was completed.There was no negative impact to the patient.A continuous flush was done.Puncture site was the groin.Kinking occurred upon removal of the inner catheter.The devices were used according to the instructions for use (ifu).The emboguard was prepared according to ifu.Additional event information received on 18-may-2026 indicated that there was no allegation of patient injury due to an alleged product issue.
 
Additional Manufacturer Narrative
Manufacturer ref# (b)(4) information regarding patient weight, height, medical history, race, and ethnicity was not reported.Since no device has been received for analysis, no product investigation can be performed and the reported failure could not be evaluated.A device history review (dhr) associated with this lot presented no issues during the manufacturing or inspection process that can be related to the reported complaint.With the information available and without the product available for analysis, the reported customer complaint could not be confirmed.Based on the device history review, there is no indication that the event is related to the device manufacturing process.The exact cause of the event could not be determined; however, there are circumstances of the procedure that may have contributed to the reported failure.The instructions for use (ifu) of the emboguard devices cautions users against advancing or withdrawing the catheter or dilator against resistance without careful assessment of the cause of resistance using fluoroscopy.If cause cannot be determined, withdraw the catheter/dilator while exercising caution.Movement against resistance may result in damage to vessel or catheter/dilator, e.G.Kinking, or cause patient injury.Additionally, if the emboguard catheter becomes severely kinked, users are instructed to withdraw the entire system, (i.E.Emboguard catheter, intermediate catheter, guidewire, etc.).Use of a damaged device may result in vessel injury.As part of the post market surveillance program, information from this complaint is trended to identify statistical signals for consideration of further correction action.Since there was no evidence to suggest the event was related to a manufacturing or design issue, no corrective actions will be taken at this time.The company is seeking this information through the event investigation.This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by cerenovus, or its employees that the report constitutes an admission that the product, cerenovus, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EMBOGUARD
Common Device Name
CATHETER, PERCUTANEOUS, NEUROVASCULATURE
Manufacturer (Section D)
NEURAVI LTD.
block 3
ballybritt business park
galway H91 K 5YD
EI  H91 K5YD
Manufacturer (Section G)
LAKE REGION MEDICAL GALWAY
parkmore business park west
county galway H91 C K22
EI   H91 CK22
Manufacturer Contact
henry boden
31 technology dr
irvine, CA 92618
9497882078
MDR Report Key25218456
Report Number3011370111-2026-00065
Device Sequence Number9778457
Product Code QJP
UDI-Device Identifier10886704082743
UDI-Public(01)10886704082743(11)251030(17)280504(10)10310876
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K212340
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 06/09/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue NumberBG8795U
Device Lot Number10310876
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 05/07/2026
Supplement Date Received by Manufacturer05/27/2026
Initial Report FDA Received Date05/19/2026
Supplement Report FDA Received Date06/09/2026
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured11/06/2025
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-