ICU MEDICAL COSTA RICA LTD. PRIMARY PLUMSET, 0.2 MICRON FILTER, CLAVE Y-SITE, PE LINED TUBING, 104 INCH; SET, ADMINISTRATION, INTRAVASCULAR
|
Back to Search Results |
|
| Catalog Number |
142550489 |
| Medical Device Problem Codes |
Crack (1135); Fluid/Blood Leak (1250)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
04/01/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
B3 - date of event - the date of event is unknown, and 01 apr 2026 has been used as a placeholder.The device has been requested for evaluation- it has not been received.The investigation is pending.
|
| |
|
Event or Problem Description
|
|
Event occurred regarding a primary plumset, 0.2 micron filter, clave y-site, pe lined tubing, 104 inch where it was reported that a couple of incidents involving a potential leak from the filter on the primary plum set occurred during hazardous drug infusions on 04-02 (paclitaxel) and 04-21 (polatuzumab).Both events involved peripheral infusions resulting in leakage of hazardous drug with exposure to both the patient and staff requiring medication remake and leading to drug wastage.The leak was associated with the filter, with one instance involving a hole while the other involved a suspected crack.One filter was confirmed cracked whereas the other showed no apparent visible damage.
|
| |
|