| Catalog Number |
00514010900 |
| Medical Device Problem Code |
Smoking (1585)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
12/01/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
This complaint is recorded by zimmer biomet under (b)(4).Review of the most recent repair record determined the peristaltic hose was off and rubbing.The peristaltic pump was replaced and resolved the reported issue.Review of the device history record identified no deviations or anomalies during manufacturing.A definitive root cause cannot be determined.The event was confirmed.If any further information is found which would change or alter any conclusions or information, a supplemental report will be filed accordingly.Zimmer biomet will continue to monitor for trends.
|
| |
|
Event or Problem Description
|
|
It was reported that outside of surgery, the device is smoking from behind.There was no patient harm/injury or surgical delay as there was no patient involvement.Due diligence is complete and there is no additional information available.
|
| |
|