• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: LEMAITRE VASCULAR, INC. TUFTEX OVER-THE-WIRE EMBOLECTOMY CATHETER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

LEMAITRE VASCULAR, INC. TUFTEX OVER-THE-WIRE EMBOLECTOMY CATHETER Back to Search Results
Catalog Number 1651-88
Medical Device Problem Codes Material Rupture (1546); Unraveled Material (1664)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 04/24/2026
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The photo provided confirmed the report balloon ruptured.While the reported event was confirmed, the exact root cause of the reported issue could not be determined.It is possible that procedural factors/anatomical features such as calcification, or plaque caused/contributed to the reported issue.The instructions for use (ifu) include the following warning regarding possible complications: in common with other catheterization procedures, complications may occur.These may include but are not limited to: infection, local hematomas, intimal disruption, arterial dissection, perforation and rupture, hemorrhage, arterial thrombosis, distal emboli of blood clots or arteriosclerotic plaque, air embolus, aneurysms, arterial spasms, arteriovenous fistula formation, balloon rupture or tip separation with fragmentation and distal embolization.Exercise caution when encountering extremely diseased vessels.Arterial rupture or balloon failure due to sharp calcified plaque may occur.The possibility of balloon rupture must be taken into account when considering the risks involved in the catheterization procedure.Due to the increased rate of occurrence of this issue, capa 2024-007 was previously implemented to document further investigation of this failure.The reported problem of ligature unraveled was also confirmed.The ligature likely slipped while pulling the device during the procedure.It could not be determined if this was due to a component failure or excessive force being used in the procedure.The production and traceability record for the device was reviewed; no issues were found during manufacturing or packaging that would cause or contribute to the reported event.All quality control tests were completed successfully and met specification.
 
Event or Problem Description
It was reported that the tuftex otw balloon ruptured and the ligature unraveled during use in a procedure.The device was checked prior to use.No patient injury was reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TUFTEX OVER-THE-WIRE EMBOLECTOMY CATHETER
Common Device Name
EMBOLECTOMY CATHETER
Manufacturer (Section D)
LEMAITRE VASCULAR, INC.
63 second ave
burlington MA 01803
Manufacturer (Section G)
LEMAITRE VASCULAR, INC
63 second ave
burlington MA 01803
Manufacturer Contact
peter song
63 second ave
burlington, MA 01803
9784251683
MDR Report Key25225081
Report Number1220948-2026-00121
Device Sequence Number13211816
Product Code DXE
UDI-Device Identifier00840663100750
UDI-Public(01)00840663100750(11)250818(17)280728(10)XOT00002912(91)1651-88
Combination Product (Y/N)N
Initial Reporter CountryKS
PMA/510(K) Number
K022145
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Company Representative
Initial Reporter Occupation Nurse
Type of Report Initial
Report Date (Section B) 05/19/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date07/28/2026
Device Catalogue Number1651-88
Device Lot NumberXOT00002912
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 04/24/2026
Initial Report FDA Received Date05/19/2026
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured08/18/2025
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-