MAUDE Adverse Event Report: ENCORE MEDICAL L.P. DJO SURGICAL; EMPOWR PARTIAL KNEETM, TIBIAL INSERT, SIZE 2, 8MM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ENCORE MEDICAL L.P. DJO SURGICAL; EMPOWR PARTIAL KNEETM, TIBIAL INSERT, SIZE 2, 8MM Back to Search Results
Catalog Number 347-08-702
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Codes Fall (1848); Pain (1994)
Date of Event 04/29/2026
Type of Reportable Event Serious Injury
Event or Problem Description
Revision surgery - due to the patient fell onto his knees and was experirncing pain in his opperative knee.
 
Additional Manufacturer Narrative
The agent reported, "patient fell onto his knees and was experiencing pain in his operative knee.After the revision surgical approach, there did not seem to be any malalignment or loosening, but it was determined that a conversion to a total knee prosthesis was the best option." the previous surgery and the surgery detailed in this event occurred 2 years, 6 months, 9 days apart.Item number: 347-08-702."item description: empowr partial kneetm, tibial insert, size 2, 8mm." lot#: 164z1018.Part revision: b.Product type: knee.Manufacture date: 29-jul-2020.Expiration date: 21-jul-2025.This evaluation is limited in scope as the item(s) associated with this investigation was not returned to djo surgical - austin for examination.The surgery was not completed as intended.If additional information regarding the reported event is submitted at a future date, this investigation will be re-evaluated.There was no information submitted with this complaint about any patient activities, accidents, or medical contraindications that may have contributed to the reported event.There were no findings during this evaluation that indicate that the reported device(s) was defective.The surgeon performed this procedure to remedy the patient's condition.No further action is deemed necessary.Root cause: the root cause of this complaint was a revision surgery due to a patient fall onto the operative knee resulting in pain, despite no evidence of malalignment or loosening, leading to conversion to a total knee prosthesis.No information was submitted with the complaint regarding the patient's pre-existing conditions or any patient activities that may have contributed to the event.There are multiple factors that may contribute to an event that are outside of the control of enovis surgical.Containment: inventory containment is not required as there are no indications of a product or process issue affecting implant safety or effectiveness.The revision surgery was completed successfully.Device history records review: a review of the device history record(s) shows that the reported component(s) used in the previous surgery met design and manufacturing requirements at the time of release for use.There were no nonconforming material reports associated with the product(s) that may have contributed to the reported event.The device(s) was verified to have gone through an acceptable sterilization process and was within its expiration date at the time of the previous surgery.Complaint history: customer complaint history of the reported device(s) showed no present trends or ongoing issues that need review.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DJO SURGICAL
Common Device Name
EMPOWR PARTIAL KNEETM, TIBIAL INSERT, SIZE 2, 8MM
Manufacturer (Section D)
ENCORE MEDICAL L.P.
9800 metric blvd.
austin TX 78758
Manufacturer (Section G)
ENCORE MEDICAL L.P.
9800 metric blvd.
austin TX 78758
Manufacturer Contact
stacey bonnell
9800 metric blvd
austin, TX 78758
5128346313
MDR Report Key25225297
Report Number1644408-2026-00975
Device Sequence Number9801798
Product Code HSX
UDI-Device Identifier00190446294870
UDI-Public00190446294870
Combination Product (Y/N)N
Initial Reporter CountryUS
PMA/510(K) Number
K191325
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2023
Device Explanted Year2026
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Company Representative
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date (Section B) 05/19/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Lay User/Patient
Device Expiration Date07/21/2025
Device Catalogue Number347-08-702
Device Lot Number164Z1018
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 04/29/2026
Initial Report FDA Received Date05/19/2026
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured07/29/2020
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
247-02-102, LOT: 160Z1038.; 357-02-102, LOT: 194Z1009.
Outcome Attributed to Adverse Event Other;
Patient Age55 YR
Patient SexMale
-
-