The product was returned to pentax medical for repair.Our technician checked the returned unit and confirmed that the ccd drive pcb fluid damage.Based on the result, we concluded that it was caused due to the fluid damage from the ccd drive pcb.In addition, our technician confirmed that the lg connector fluid damage, the lcb (light carrying bundle) broken, the lg connector corroded, the shield cover corroded, the air/water socket deformed, the bending rubber leak, the operation channel (primary) cut, the bending rubber cut, the distal body worn out, the insertion flexible tube bump, and the suction channel kink; however, these defects are not the main cause, and/or irrelevant to the alleged complaint.Based on the technical report "hr-rpt-0586(image failure)" and/or the risk analysis results, it was evaluated to submit mdr.
|