|
The product was returned to pentax medical for repair.Our technician checked the returned unit and confirmed that the bending rubber gluing missing.Based on the result, we concluded that it was caused due to the physical damage applied on the bending rubber gluing.In addition, our technician confirmed that the ultrasound connector body broken, the remote control buttons leak, the light guide cable leak, the remote control buttons cut, the light guide cable cut, and the us connector cable (long) bump; however, these defects are not the main cause, and/or irrelevant to the alleged complaint.In terms of the bending rubber glue missing, the possibility of dropping into human body could not be denied.Moreover, based on the technical report "hr-rpt-0581(bending rubber)" and/or the risk analysis results, it was evaluated to submit mdr.
|