On (b)(6) 2026, a patient underwent a ultrasound guided kidney biopsy procedure through normal density tissue using maxcore instrument.It was reported that during the procedure, the instrument outer cannula allegedly could not be fired.No coaxial needle was used.The procedure was completed using another device.There was no reported patient injury.
|
|
H11: manufacturing review: the device history records have been reviewed, and this lot met all release criteria.Investigation summary: the physical device was not returned for evaluation.Three electronic photos were provided and reviewed.The first photo shows that a labeling information of a maxcore device which did match and verifies with tw details.The second photo shows the conversation chat; one picture is related to labeling information remaining one shows that six maxcore device were at half primed position.The third photo shows that half of the maxcore device in initial position.No other anomalies are noted.Therefore, the investigation is inconclusive for the reported failure as no objective evidence was provided for review.A definitive root cause for the reported failure to fire issue could not be determined based upon the provided information.It is unknown whether patient and/or procedural issues contributed to the reported event.Labeling review: as the reported event did not allege a labeling or use related issue, a labeling review is not required section a through f: the information provide by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant/reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
|